产品名称:JK9711 Electronic load
规格:
产品备注:
产品类别:Electronic load
Product Description:       
 JK9711 series electronic load is our company on the basis of extensive user feedback and collection of electronic load years of professional accumulation.The newly designed circuit uses fast AD and DA to ensure high-speed and high-precision measurement. The resolution of 0.1mV/0.1mA can more effectively monitor the detailed changes of voltage and current, so that it can be applied to a wider range of testing better results.Highlight VFD makes the display clear and intuitive, menu and display content continue the original electronic load series easy to understand and operate simple style, easy to use.Key band operation status indicator.The working state is clear at a glance.The complete test function is highly humanized and greatly improves production efficiency. The RS232 interface protocol is fully compatible with the original series of electronic loads, making the system updating no longer cumbersome.
The new CR-LED test mode can simulate the characteristics of the LED lamp. By adding the on-voltage setting of the diode, the working principle of the diode can be simulated completely. The test voltage and current can reach a normal stable value, which avoids the instability of voltage and current in the traditional constant resistance mode. It reflects the actual carrying condition of LED drive power supply.
The new generation of electronic load can meet the LED drive power supply, switching power supply, power transformer, charger, battery and other industries in the production line test, product development and other experimental research and development areas.
Main performance characteristics:
The CR-LED measurement mode completely simulates the load characteristics of LED power supply.       
■High brightness, visible vacuum fluorescent VFD display, clear and comprehensive.
■Voltage 0.1mV, current 0.1mA high display resolution.
■Function of OCP and OPP to accurately capture critical parameters.
■A complete CC, CV, CR, CP mode
■Battery test mode can automatically record discharge time and capacity.
■The dynamic conversion time of 0.1mA can effectively investigate the dynamic response of the power source.
■Listing test can flexibly combine test mode and time to determine test results.
■Remote measurement function, multi group data storage
■Intelligent fan control and various protection measures                           
 

Technical indicators:

Model  JK9711 JK9712 JK9712A JK9711A JK9712B JK9711B
Rated value Input
voltage
0~120V 0~500V 0~60V
Input current 1mA~30A 1mA~30A 1mA~60A
Input power 150W 300W 300W 150W 300W 150W
  range  accuracy resol-
ution
accuracy resol-
ution
accuracy resolu-
tion
Load precision 0-18V ±(0.05%+
0.03%FS
1mV ±(0.05%+
0.03%FS)
1mV ±(0.05%+
0.03%FS)
1mV
0-120V/
500V
±(0.05%+
0.03%FS)
10mV ±(0.05%+
0.03%FS)
10 mV ±(0.05%+
0.03%FS)
10mV
  0-3A ±(0.05%+
0.05%FS)
0.1mA ±(0.05%+
0.05%FS)
0.1 mA ±(0.05%+
0.05%FS)
0.1mA
0-30A/
60A
±(0.05%+
0.05%FS)
1 mA ±(0.05%+
0.05%FS)
1 mA ±(0.05%+
0.05%FS)
1 mA
Constant voltage mode 1.5V-18V ±(0.05%+
0.03%FS)
1mV ±(0.05%+
0.03%FS)
1mV ±(0.05%+
0.03%FS)
1mV
1.5V-120V/
500V
±(0.05%+
0.03%FS)
10mV ±(0.05%+
0.03%FS)
10mV ±(0.05%+
0.03%FS)
10mV
Constant current mode 0-3A ±(0.05%+
0.05%FS)
0.1mA ±(0.05%+
0.05%FS)
0.1 mA ±(0.05%+
0.05%FS)
0.1 mA
0-30A
/60A
±(0.05%+
0.05%FS)
1mA ±(0.05%+
0.05%FS)
1mA ±(0.05%+
0.05%FS)
1mA
Constant resistance mode
(when input voltage and current value)≥
10%FS)
0.05Ω-
±(0.2%+
0.2%FS)
0.001Ω ±(0.2%
+0.2%FS)
0.001Ω ±(0.2%+
0.2%FS)
0.001Ω
0.5Ω-
50Ω
±(0.1%+
0.1%FS)
0.01Ω ±(0.1%+
0.1%FS)
0.01Ω ±(0.1%+
0.1%FS)
0.01Ω
5Ω-
500Ω
±(0.1%+
0.1%FS)
0.1Ω ±(0.1%+
0.1%FS)
0.1Ω ±(0.1%+
0.1%FS)
0.1Ω
  500Ω-
5KΩ
±(1%+
1%FS)
±(1%+
1%FS)
±(1%+
1%FS)
Constant power mode
(when input voltage and current value)≥
10%FS)
0-50W ±(0.1%
+0.1%FS)
1mW ±(0.1%+
0.1%FS)
1 mW ±(0.1%+
0.1%FS)
1 mW
0-150W ±(0.1%+
0.1%FS)
1 0mW ±(0.1%+
0.1%FS)
10 mW ±(0.1%+
0.1%FS)
10 mW
  0-300W ±(0.1%+
0.1%FS)
0.1W ±(0.1%+
0.1%FS)
0.1 W ±(0.1%+
0.1%FS)
0.1 W
Voltage measurement accuracy 0-9.9999V ±(0.05%+
0.03%FS)
0.1mV ±(0.05%+
0.03%FS)
0.1mV ±(0.05%+
0.03%FS)
0.1mV
10.000-99.999V ±(0.05%+
0.03%FS)
1mV ±(0.05%+
0.03%FS)
1mV ±(0.05%+
0.03%FS)
1mV
  100.00-120.00V/
500.00V
±(0.05%+
0.03%FS)
10mV ±(0.05%+
0.03%FS)
10mV ±(0.05%+
0.03%FS)
10mV
Current measurement accuracy 0-9.9999A ±(0.05%+
0.05%FS)
0.1mA ±(0.05%+
0.05%FS)
0.1 mA ±(0.05%+
0.05%FS)
0.1 mA
10.000-30.000/
60.000A
±(0.05%
+0.05%FS)
1 mA ±(0.05%
+0.05%FS)
1mA ±(0.05%+
0.05%FS)
1mA
Battery test function                       Input= 0.8-120V/500V         Max measurement capacity= 999A/H
Resolution=0.1mA       Timer range=1~60000sec                                                      
Dynamic testing mode T1&T2(A or B test time):0.1mS-999S  Error<2.5% + 0.1mS                          
Scope of protection >specified condition 5%
Input impedance ≥200KΩ
size W*H*D230mm*100mm*350mm
weight  6.5Kg

The main test functions of the load are:
 
Constant current mode:
In the constant current mode, the electronic load is simulated as a constant current load, and the current is within the range of the output capacity of the input power supply and does not change with the voltage.
In order to reduce the impulse effect at the instant of current starting, the rise time can be set and the current can be slowly raised by load control during this time.
In the process of constant current, it can be directly switched to the short-circuit working state. Load control works with maximum current to test the short-circuit output and protection capacity of power supply.
 
Constant voltage mode:
In the constant voltage mode, the electronic load is simulated as a constant voltage load, and the output voltage does not change with the load current within the output capacity of the power supply.
In the process of constant voltage, it can be directly switched to the short-circuit working state. Load control works at the minimum voltage to test the short-circuit output and protection capacity of the power supply.
Constant power mode:
In the constant power mode, the electronic load consumes a constant power in the range of the power output capacity. It is generally simulated in a constant current mode, and the load current is dynamically adjusted according to the input voltage. In constant power mode (CP), the load consumes a constant power. When the input voltage changes, the load will adjust the current to maintain the power consumed.
Constant resistance mode:
In constant resistance mode, the load works in constant current or voltage mode. According to the change of input power supply, the load current or voltage is dynamically adjusted to make the load equivalent to a constant resistance. In constant resistance mode (CR), the load is equivalent to a constant resistance, and the load consumes the current that changes with the input voltage.
Battery test mode:
Discharge test works in a constant current (CC) or constant resistance (CR) mode, and can be automatically tested in the same mode or different modes according to the set conversion time to better simulate the actual working state of the battery; during continuous discharge, the battery voltage drops continuously, when the load input voltage is lower than the set value, the discharge test. Test automatically stop, the instrument can display discharge time and battery capacity.
 
Dynamic measurement mode:
Dynamic testing allows the load to be repeatedly switched between two voltages or currents, which can be used to test the dynamic characteristics of the power supply. After the start of the dynamic test, the load will continuously switch between the A value and the B value, maintaining the width A and the width B respectively. Continuous measurement (CONT), pulse measurement (PULS) and trigger measurement (TRIG) can be set up.
 
List automatic test mode:
The list test function can be automatically converted at set time under different load modes. For power supply products and charger equipment, through multi-parameter hybrid testing, can be more comprehensive and in-depth understanding of the product under test in various applications under the working characteristics.

 

Programming test:

Programming test can test some critical parameters of power supply products, such as current limiting protection current, turning point current, etc.By setting the scanning range and critical conditions, the electronic load can capture the voltage, current or power parameters of the power supply before the critical conditions occur, and can set the comparator to judge whether the power supply is qualified or not

 

Remote test function:

When the load consumes a large amount of current, the voltage drop will occur on the connecting line between the power source under test and the load, thus affecting the voltage measurement accuracy. In CV, CR, CP mode, voltage sampling accuracy will affect the working accuracy of the electronic load, the use of remote testing function can completely eliminate the impact of voltage drop on the test.

 

Standard RS232C interface:

The instrument has a standard RS232 interface. Users can communicate with the instrument by using the general PC software provided by the company or by their own programming.

We are professional supplier of instrument! www.jk17.com is our communication platform
Submit orders
All information in the confirmation page after the check, please click "submit" at the bottom of the page, if enter the settlement of a complete page shows your checkout process has ended, order has generated, in the settlement inflammation of the order page you can also understand the culture order number (it is recommended that you will be inflammation from the order number and customer order number recorded at the same time, in order to "order inquiry"). Instrument mall management system will be your order send Email to confirm your order information. Reply to order by received payment authorization, JK17 instrument network will complete delivery and accounts processing. Do order at any time if you want to know, you can query in the "order inquiry" function of your order. JK17 instrument ? network also prepared fax form of order
Goods payment for goods
Customers in Shanghai in order after check, please wait for our delivery. Customers outside Shanghai, changzhou, please order after check to wire payment or deposit to us, see remittance way remittance way... more>>
Tel :(+86)4001128155 0519-8556347,89187775
0755-61282381,27467016
Email: mailjk17@126.com
copy Changzhou Jinailian Electronic Technology Co., Ltd 2009 All Rights Reserved.
Technical support:WWW.WIN-ALL.CN     Su ICP preparation 06033252